- Model: Infusion Pump EPSIMED ZNB-XAII
- Pumping Mechanism Curvilinear peristaltic
- IV Set Compatible with IV sets of any standard
- Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)
- Purge, Bolus 100-1500 ml/h (in 0.1 m/h increments)
- Purge when pump stops, bolus when pump starts,
- Accuracy ±3%
- VTBI 1-20000 ml (in 0.1 ml increments)
- Infusion Mode ml/h, drop/min, time-based, body weight, nutrition
- KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)
- Alarms Occlusion, air-in-line, door open, end program, low
- battery, end battery, AC power off, motor malfunction,
- system malfunction, reminder alarm
- Additional Features Real-time infused volume, automatic power switching,
- mute key, purge, bolus, system memory, history log,
- drug library, key locker, rotary knob, change flow rate
- without stopping the pump
- Occlusion Sensitivity High, medium, low
- History Log 50000 events
- Air-in-line Detection Ultrasonic detector
- Wireless management Optional
- Drop Sensor Optional
- Vehicle Power (Ambulance) 12±1.2 V
- Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA
- Battery 9.6±1.6 V, rechargeable
- Battery Life 5 hours at 25 ml/h
- Working Temperature 10-30 degrees
- Relative Humidity 30-75%
- Atmospheric Pressure 860-1060 hpa
- Size 130*145*228 mm
- Weight 2.5 kg
- Safety Classification Class.